A&D Counting Scale HC-15Ki User Manual
Page 51
49
Unit Weight
Command ?
U W
C L
F
Reply U
W ,
+
1
.
2 3 4 5 6 7
[
[ g C L
F
U
W ,
+
0
.
2 7 2 5 3 1
[
l
b C L
F
Total Count
Command ?
A Q
C L
F
Reply
A Q ,
+ 0 0 9 9 9 9 9 9
[ P C C L
F
Accumulation Numbers
Command ?
A N
C L
F
Reply
A N ,
0 0 0 0 1 2 3 4 C L
F
Tare Weight
Command ?
T R
C L
F
Reply
T R ,
+ 0 0 1
.
2 3 4 6
[ k g C L
F
Request the scale to reply with the contents of ID memory.
Command ?
M R , 1
2
C L
F
Request the contents of id-12.
Reply
M R ,
0 0 0 0 1 2
,
1 2
.
3 4 5 6 7
,
+ 0 0 0
.
2 3 4 5
,
+ 0 0 0 0 1 3 5 7
,
+ 0 0 0 0 1 2
4 6
,
+ 0 1 2 3 4 5 6 7
,
0 0 0 0 1 2 3 4 C L
F
MR, ID number (6 digit), unit weight (8 digit including decimal point), tare weight (9
digit including sign and decimal point), upper limit (9 digit including sign), lower
limit (9 digit including sign), total count (9 digit including sign), number of addition
(8 digit)
C
R
L
F
.
Memorize (enter) unit weight and tare weight into a specified ID memory.
Command M R , 1 2 , 1
.
2 3
,
0
.
3 4 5 C L
F
Reply ACK C L
F
ID Number:
Maximum 6 digit
Unit Weight:
Maximum 8 digit including decimal point
Tare Weight:
Maximum 8 digit including decimal point (‘kg’ or ‘lb’ depends on
f-00 setting).
Store the comparator limits into a specified ID memory.
Command M L
, 1 2 , + 1 0 1 0
,
+ 9 9 0 C L
F
ID # Unit Weight Tare Weight
ID # Upper Limit Lowe Limit