beautypg.com

A&D Counting Scale HC-15Ki User Manual

Page 47

background image

45

Baud Rate (f-06-04)

Select the baud rate according to the device to be connected.

‰ 2400 bps (f-06-04="0") Select 2400 bps to connect with an AD-8121.

‰ 4800 bps (f-06-04="1")

‰ 9600 bps (f-06-04="2")

13-4. Connecting the AD-8121 Printer / MODE 1 or MODE 2

‰ When using the AD-8121 printer (MODE 1 or MODE 2), you will be able to get data:

Number of data items, total, maximum, minimum, mean value, range of data (max. -
min. data) and standard deviation.

‰ When using the AD-8121 with MODE 2, set f-06-02 to print pcs (count) data only

or weight data only.

‰ To print date and time, use the AD-8121’s calendar / clock function and set f-06-02

to print pcs (count) data only or weight data only.

Print Operations Settings

Print By:

F-Function f-06-01 Printer

MODE

HC PRINT key

0

MODE 1

Auto Print

2 or 3 MODE

1

Printer DATA key

1

MODE 2

Example of f-06-02 settings

‰ To print pcs (count) data only: set f-06-02 at "0100"

‰ To print weight data only: set f-06-02 at "0010"

‰ To print pcs (count) and weight data: set f-06-02 at "0110"

‰ To print pcs, weight and unit weight data: set f-06-02 at "0111"

‰ To print total data (accumulated by the M+ key), press the TOTAL key so the

count display shows the total, then press the PRINT key.

‰ If you are using the AD-8121’s statistic functions, then set f-06-02 at "01#0" (# = 0

or 1) for pcs (count) data or "0010" for weight data.

‰ MODE 1 and 2 of the AD-8121 can not print ID numbers.