88 cal 0 f 8, 11hi00 – A&D Counting Scale HC-15Ki User Manual
Page 30
28
If you know the exact weight value, or
if you wish to use a different weight,
use the 0
Æ 9 and . 10-key pad to
display the desired calibration weight.
(For example: Using 5 kg calibration weight
that actually weighs 5.001 kg. Do not forget
to enter a decimal point in this case.)
3. Press the ENTER key.
The calibration weight stops blinking.
4. Making sure that there is nothing on, or touching
the weighing pan, press the ENTER key.
When zero calibration is completed, the display
will show “Cal f”.
If you don’t need span calibration, press the
ON/OFF key to exit from the calibration
procedure.
5. Place the calibration weight on the weighing pan
and press the ENTER key.
When span calibration is completed, the display
returns to step 1 showing the weight value for the
calibration weight.
Remove the calibration weight.
If the calibration weight is not what it should be,
an error will be displayed. Check if the weight is
correct and try again.
6. Press the ON/OFF key to turn the scale off
and re-attach the calibration switch cover.
(End of the calibration procedure.)
If the scale will be moved to another place, set the gravity
acceleration value before calibration. The value must be of the
area where the calibration is to be done.
888
Cal
10
COUNT
WEIGHT UNIT WEIGHT
88
Cal
0
0
8
8
0.0
5.001
0
00
888
Cal
10
COUNT
WEIGHT UNIT WEIGHT
88
Cal
0
0
8
8
0.0
5.000
0
00
ENTER
888
Cal
10
COUNT
WEIGHT UNIT WEIGHT
88
Cal
0
f
8
8
0.0
5.000
0
00
Nothing on the
weighing pan!
ENTER
888
Cal
02
COUNT
WEIGHT UNIT WEIGHT
88-
5.000
8
8
11Hi00
Place
Calibration
i ht
ENTER