‰ ID Memory contents.

The scale ID memory can contain unit weights with tare
weights and comparator limits, or just unit weights alone.

00

ID memory contains the unit weight only.

You select which data to be stored by keying in a

0 or 1 for the data: tare weight or comparator

limits.
Example: Key in 1

0 to display 10, ID memory

contains the unit weight and comparator limits.

F-02-X ACAI Operation & Min. Unit Weight

‰ ACAI Mode When Unit Weight entered by Sample Pieces.

0

ACAI is disabled.

1

ACAI automatic Operation.

2

ACAI Manual Mode (using the ENTER key).

‰ ACAI Mode When Unit Weight entered by Keyboard or ID.

0

ACAI is disabled.

1

ACAI Manual Mode (using the ENTER key). This setting
works when f-02-01 setting is not “0”.

2

ACAI Automatic obeys f-02-01 setting.

‰ Minimum Unit Weight. (1d=1 weighing display division)

The factory setting may be different for some countries.

0

1/5 d

1

1/100 d

f-02-01

f-02-02

f-02-03

f-01-05

00

Comparator limits