beautypg.com

11hi00 – A&D Counting Scale HC-15Ki User Manual

Page 31

background image

29

8-2. Gravity Compensation

When the scale is first used or has been moved to different place, it should be
calibrated using a calibration weight.
But if the calibration weight cannot be prepared, the gravity acceleration correction
will compensate the scale. Change the gravity acceleration value of the scale to the
value of the area where it will be used. Refer to the gravity acceleration map
appended to the end of this manual.

1. In the Step 1 above, press TARE key

The display flashes the gravity acceleration
value stored in the scale.

2. Use the 0

Æ

9 10-key pad to

display the desired gravity
acceleration value.
(Example of the value 9.800 m/s

2

.)

3. Press the ENTER key.

The scale will store the new value.
If necessary to calibrate the scale
using a weight, go to Step 2 of the
previous section.

4. Press the ON/OFF key to turn the scale off and re-attach the calibration switch

cover. (End of the calibration procedure.)

ENTER

888

Cal

10

COUNT

WEIGHT UNIT WEIGHT

88

g

al008

8

0.0

9.798

0

00

888

Cal

10

COUNT

WEIGHT UNIT WEIGHT

88

g

al008

8

0.0

9.800

0

00

888

Cal

02

COUNT

WEIGHT UNIT WEIGHT

88-

0.000

8

8

11Hi00

This manual is related to the following products: