Real time screen – Campbell Scientific TGA100 Trace Gas Analyzer Manual User Manual
Page 28

The functions available at the main menu are described below.
R) Real Time TGA Program
Turns the TGA on and displays the real time screen. This is the normal operating mode. See section 3.4 for
additional information.
T) TGA
on/off
Toggles the TGA on or off. When the TGA is on, all current and temperature controls are active and
concentration calculations are being made, but the real time screen is not displayed, and no data are saved
to the hard disk.
L) Laser
on/off
Toggles the laser on or off. The laser must be on during normal operation. The laser may be disabled to
operate the TGA100 without driving the laser. For example, the laser temperature may be monitored
during the initial cool down by disabling the laser at the main menu and then entering the real time screen.
P) Parameter Change Menu
Displays submenus for changing parameters. See section 3.5 for additional information.
M) Laser Mapping Menu
Displays the mapping submenu to be used to characterize a laser. See section 4.3 for additional information.
X) Exit
Exit TGA program.
3.4 Real Time Screen
The Real Time Screen is entered by pressing “R” at the main menu (see section 0.) This is the normal operating mode
for the TGA100. Concentration data can be displayed or saved only while in the real time mode.
28