Appendix A. Options for File Save and Real Time Display

Display Option Description

Notes

10 Hz Conc

Trace gas concentration, in ppm.

1

Mean Conc

Mean concentration measurements (ppm). If in site means or gradient mode, the mean
concentration is calculated at each valve switch. Otherwise the averaging period is specified
by dynamic parameter #12 (see section 3.4.4).

1

Conc StDev

Standard deviation of the concentration (ppb), calculated over the period specified by
dynamic parameter #12. (see section 3.4.4).

1

Laser Temp

Temperature (K) of the laser mounting plate.

Laser Heater

Voltage applied to the heater on the laser mounting plate to maintain the laser at the
specified temperature.

DC Current

Laser DC current (mA)

Pressure

Pressure in the analyzer vacuum manifold.

The units are selected by the user in the Pressure Calculation screen.

Valve Status

Status of the 16 digital IO ports used to control valves in the Site Means or Gradient mode.
The valve status is reported as a number from 0 to 65535.

Ref Signal

Reference detector signal at the center of the spectral scan, (mV), corrected for gain and
offset.

1

Ref Trans

Reference detector transmittance at the center of the spectral scan (%)

1

Ref Temp

Reference detector temperature (°C)

Ref Peltier

DAC command to supply current to the Peltier cooler to maintain the reference detector at
the specified temperature

Ref Gain

Reference detector gain

Ref Offset

Reference detector offset

Smp Signal

Sample detector signal at the center of the spectral scan, (mV), corrected for gain and offset

1

Smp Trans

Sample detector transmittance at the center of the spectral scan (%)

1

Smp Temp

Sample detector temperature (°C)

Smp Peltier

DAC command to supply current to the Peltier cooler to maintain the sample detector at the
specified temperature

Smp Gain

Sample detector gain

Smp Offset

Sample detector offset

Ref Signal B

Reference detector signal at the center of the spectral scan, (mV), corrected for gain and
offset, for ramp B

2

Ref Trans B

Reference detector transmittance at the center of the spectral scan (%), for ramp B

2

Smp Signal B

Sample detector signal at the center of the spectral scan, (mV), corrected for gain and offset,
for ramp B

2

Smp Trans B

Sample detector transmittance at the center of the spectral scan (%), for ramp B

2

10 Hz Conc B

Trace gas concentration (ppm), for ramp B

2

Mean Conc B

Mean concentration (ppm) for ramp B. If in site means or gradient mode, the mean
concentration is calculated at each valve switch. Otherwise the averaging period is specified
by dynamic parameter #12 (see section 3.4.4).

2

Conc StDev B

Standard deviation of the concentration (ppb), calculated over the period specified by 2

104