Appendix B. SDI-12 Sensor Support

8. To change the SDI-12 address, enter the command aAb!; where a is the

current address from the above step and b is the new address. The CS650

will change its address and the datalogger will exit the SDI-12 Transparent

Mode.

9. Activate the SDI-12 Transparent Mode on Control Port 1 again by entering

*8#1 . Verify the new SDI-12 address by entering the ?! command.

The CS650 will respond with the new address.

10. To exit the SDI-12 Transparent Mode, type in

*0.

FIGURE B-4. SDI-12 transparent mode on CR10X-PB table-based

datalogger using control port 1 and changing SDI-12 address from

0 to 1