beautypg.com

3 operational specifications, Operational specifications, 1. cs650 and cs655 average current drain – Campbell Scientific CS650 and CS655 Water Content Reflectometers User Manual

Page 13

background image

CS650 and CS655 Water Content Reflectometers

FIGURE 5-1. CS650 and CS655 average current drain

FIGURE 5-1 shows average current drain for different measurement rates and

quantities of CS650 probes. If the time between measurements is five minutes

or longer, average current drain may be approximated at 0.15 milliamps per

sensor.

5.3 Operational Specifications

CS650

CS655

Relative Dielectric

Permittivity

Range:

1 to 81

1 to 81

Accuracy†:

1 to 40:

40 to 80:

±

(2% of reading +

0.6) for solution EC

3 dS/m

±

1.4 for solution EC

3 dS/m

±

(3% of reading +

0.8) for solution EC

8 dS/m

±

2 for solution EC

2.8 dS/m

Precision‡:

<0.02

<0.02

5