beautypg.com

Campbell Scientific CS650 and CS655 Water Content Reflectometers User Manual

Page 14

background image

CS650 and CS655 Water Content Reflectometers

CS650

CS655

Volumetric Water

Content using Topp

Equation (m

3

/m

3

)

Range:

5% to 50%

5% to 50%

Accuracy†:

±

3% VWC typical in

mineral soils where

solution EC ≤3 dS/m

±

3% VWC typical in

mineral soils where

solution EC ≤10 dS/m

Precision‡:

<0.05%

<0.05%

Electrical

Conductivity

Range Solution

EC:

0 to 3 dS/m

0 to 8 dS/m

Range Bulk EC: 0 to 3 dS/m

0 to 8 dS/m

Accuracy†:

±

(5% of reading +

0.05 dS/m)

±

(5% of reading +

0.05 dS/m)

Precision‡:

0.5% of BEC

0.5% of BEC

Temperature

Soil

Measurement

Range:

-10 to + 70°C

-10 to + 70°C

Operational

Range:

0 to + 70°C

0 to + 70°C

Accuracy†:

±

0.5°C for probe

body buried in soil

±

0.5°C for probe

body buried in soil

Precision‡:

±

0.02°C

±

0.02°C

Sensing Volume*:

7800 cm

3

3600 cm

3

*

Sensing Volume approximately 7.5 cm radius around each probe rod

and 4.5 cm beyond the end of the rods
†Accuracy specifications are based on laboratory measurements in a

series of solutions with dielectric permittivities ranging from 1 to 81

and solution electrical conductivities ranging from 0 to 3 dS/m.
‡Precision describes the repeatability of a measurement. It is

determined for the CS650 by taking repeated measurements in the

same material. The precision of the CS650 is better than 0.05 %

volumetric water content and 0.01 dS/m electrical conductivity.

6