beautypg.com

Campbell Scientific CS650 and CS655 Water Content Reflectometers User Manual

Page 38

background image

CS650 and CS655 Water Content Reflectometers

The container to hold the soil during calibration should be non-metal and large

enough that the rods of the probe are no closer than about 10 cm from any

container surface.

Pack the container as uniformly as possible in bulk density with relatively dry

soil (volumetric water content <10%).

Probe rods can be buried in a tray or inserted into a column. When using a

column, insert the rods carefully through surface until rods are completely

surrounded by soil. Movement of rods from side-to-side during insertion can

form air voids around rod surface and lead to measurement error.

Collect the probe permittivity output. Repeat previous step and this step 3 or 4

times.

Determine volumetric water content by subsampling soil column after

removing probe or using mass of column. If subsampling is used, remove soil

from column and remix with samples used for water content measurement.

Repack column.

Water can then be added to the top of the container. It must be allowed to

equilibrate. Cover the container during equilibration to prevent evaporation.

The time required for equilibration depends on the amount of water added and

the hydraulic properties of the soil. Equilibration can be verified by frequently

observing the CS650 permittivity output. When permittivity is constant,

equilibration is achieved. Collect a set of calibration data values and repeat

the water addition procedure again if needed.

With soil at equilibrium, record the CS650 permittivity.

Take subsamples of the soil using containers of known volume. This is

necessary for measurement of bulk density. Copper tubing of diameter ≥ 1”

and length about 2” works well. The tubes can be pressed into the soil surface.

It is good to take replicate samples. Three carefully handled samples will

provide good results.

The sample tubes should be pushed evenly into the soil. Remove the tube and

sample and gently trim the ends of excess soil. Remove excess soil from

outside of tube.

Remove all the soil from tube to a tray or container of known mass that can be

put in oven or microwave. Weigh and record the wet soil mass.

Water is removed from the sample by heating with oven or microwave. Oven

drying requires 24 hours at 105°C. Microwave drying typically takes 20

minutes depending on microwave power and sample water content. ASTM

Method D4643-93 requires heating in microwave for 3 minutes, cooling in

desiccator then weighing and repeating this process until measured mass is

constant.

30