Appendix B. SDI-12 Sensor Support

Change Address Command (aAb!)

Sensor address is changed with command aAb!, where a is the current address

and b is the new address. For example, to change an address from 0 to 2, the

command is 0A2! The sensor responds with the new address b, which in this

case is 2.

Send Identification Command (aI!)

Sensor identifiers are requested by issuing command aI!. The reply is defined

by the sensor manufacturer, but usually includes the sensor address, SDI-12

version, manufacturer's name, and sensor model information. Serial number or

other sensor specific information may also be included.

An example of a response from the aI! command is:

313CampbellCS65X 000Std.00.35=2196405

where:

Address = 3

SDI-12 version =1.3

Manufacturer = Campbell

Sensor model = CS65X

OS version = 000Std.00.35

Sensor serial number = 2196405

Start Measurement Commands (aM! )

A measurement is initiated with M! commands. The response to each

command has the form atttnn, where

a = sensor address

ttt = time, in seconds, until measurement data are available

nn = the number of values to be returned when one or more subsequent D!

commands are issued.

Start Measurement Command (aMv!)

Qualifier v is a variable between 1 and 3 that requests variant data. Variants

include different subsets of the CS650 probe output:

M0!

Volumetric Water Content (θ), Bulk Electrical Conductivity

(σ), Temperature (°C)

M1!

Permittivity (ε), Bulk Electrical Conductivity (σ), Temperature (°C)

M2!

Period (τ), Voltage Ratio (α), Temperature (°C)

M3!

Volumetric Water Content (θ), Bulk Electrical Conductivity

(σ), Temperature (°C), Permittivity (ε), Period (τ), Voltage Ratio (α)

B-2