Troubleshooting – Campbell Scientific CS650 and CS655 Water Content Reflectometers User Manual

Page 42

background image

CS650 and CS655 Water Content Reflectometers

10. Troubleshooting

Symptom

Possible Cause

Solution

All CS650 output

values read 0

No SDI12Recorder

instruction in

datalogger program

Conditional

statement that

triggers reading is not

evaluating as true

Add SDI12Recorder

instruction (P105 for Edlog

dataloggers) to datalogger

program

Check logic of conditional

statement that triggers

readings

First value reads

NAN and all other

values read 0* or

never change from

one measurement to

another

CS650 SDI-12

address does not

match address

specified in

datalogger program

Change probe address or

modify program so that

they match

(*or all values read

NAN if the program

examples in this

manual are followed)

CS650 green wire not

attached to SDI port

specified in

datalogger program

Connect wire to correct

control port or modify

program to match wiring

CS650 not being

powered

SW12V channel not

turning on

Make sure red wire is

connected to 12V or

SW12V and black wire to

G.

If using SW12 to power

sensor, make sure red wire

is connected and

datalogger program

switches SW12 on.

If using SW 12V on a

CR10X, ensure that a wire

connects a control port to

SW12V CTRL and the

program sets that control

port high.

VWC reading is

9999999

Soil bulk permittivity

is outside probe’s

operational range

Modify program to collect

permittivity value and try

soil specific calibration

EC reading is

9999999

Soil bulk electrical

conductivity is

outside probe’s

operational range

If using CS650, try CS655

Readings erratic,

including NAN’s and

9999999’s

Multiple probes with

same SDI-12 address

sharing same control

port

Give probes unique

addresses or put on

separate control ports

34