SAFETY SUMMARY

All safety related regulations, local codes and instructions that appear in the

manual or on equipment must be observed to ensure personal safety and to

prevent damage to either the instrument or equipment connected to it. If

equipment is used in a manner not specified by the manufacturer, the protection

provided by the equipment may be impaired.

Do not use the controller to directly command motors, valves, or other

actuators not equipped with safeguards. To do so can be potentially harmful to

persons or equipment in the event of a fault to the controller. An independent

and redundant temperature limit indicator with alarm outputs is strongly

recommended.

SPECIFICATIONS

1. POWER: Power will be supplied by the Graphite host device.

2. LEDs*:

STS - Status LED shows module condition

OP1, OP2, OP3, OP4 - Indicate status of outputs 1, 2, 3, and 4

ALM, or AL1 and AL2 - Alarm LEDs are lit during any internal alarm

condition

* Default configuration.

3. MEMORY: Non-volatile memory retains all programmable parameters.

4. INPUT:

GENERAL:

Sample Time: 50 msec (20 Hz)

Common Mode Rejection: >110 dB, 50/60 Hz

Normal Mode Rejection: >50 dB, 50/60 Hz

Temperature Coefficient: 0.01%/°C

Step Response Time: 100 msec typ., 200 msec max

THERMOCOUPLE INPUTS:

Types: T, E, J, K, R, S, B, N, C

Input Impedance: 20 M Ω

Lead Resistance Effect: 0.25 µV/Ω

Cold Junction Compensation: Less than ±1°C typical (±1.5°C max) over

-40 to 75 °C ambient temperature

Resolution: 0.1°

TYPE

MEASUREMENT

RANGE

WIRE COLOR

ANSI

BS 1843

T

-200 to +400°C

-328 to +752°F

(+) Blue

(-) Red

(+) White

(-) Blue

E

-200 to +730°C

-328 to +1346°F

(+) Violet

(-) Red

(+) Brown

(-) Blue

J

-200 to +760°C

-328 to +1400°F

(+) White

(-) Red

(+) Yellow

(-) Blue

K

-200 to +1350°C

-328 to +2462°F

(+) Yellow

(-) Red

(+) Brown

(-) Blue

R

0 to +1768°C

+32 to +3214°F

No Standard

(+) White

(-) Blue

S

0 to +1768°C

+32 to +3214°F

No Standard

(+) White

(-) Blue

B

+149 to +1820°C

+300 to +3308°F

No Standard

No Standard

N

-200 to +1300°C

-328 to +2372°F

(+) Orange

(-) Red

(+) Orange

(-) Blue

C

W5/W6

0 to +2315°C

+32 to +4199°F

No Standard

No Standard

mV

0 mV to 50 mV

N/A

N/A

RTD INPUTS:

Type: 2 or 3 wire

Excitation: 150 µA

Lead Resistance: 15 Ω Max

Resolution: 1 or 0.1°

TYPE

INPUT TYPE

RANGE

385

100 Ω platinum, Alpha = .00385

-200 to +600°C

-328 to +1100°F

392

100 Ω platinum, Alpha = .003919

-200 to +600°C

-328 to +1100°F

672

120 Ω nickel, Alpha = .00672

-80 to +215°C

-112 to +419°F

PROCESS INPUT:

INPUT

RANGE

ACCURACY

IMPEDANCE

MAX CONTINUOUS

OVERLOAD

RESOLUTION

10 V

0.1% span

1 M Ohm

50 V

16 bit

20 mA

0.1% span

10 Ohm

100 mA

16 bit

5. PROBE BREAK RESPONSE: Upscale drive, Input Fault Alarm bit set high,

ALx LED illuminates.

6. TEMPERATURE INDICATION ACCURACY: ± (0.3% of span, +1°C).

Includes NIST conformity, cold junction effect, A/D conversion errors,

temperature coefficient and linearization conformity at 23 °C after 20 minute

warm up.

7. ISOLATION LEVEL: 500 Vrms @ 50/60 Hz for 1 minute between the

following:

OP1

OP2

OP3

OP4

Linear Output (GMP1 only)

Signal Input 1 and HCM

Signal Input 2 and HCM (GMP2 only)

Power Supply Input

8. COMMUNICATIONS: Provided by the graphite host device

9. A/D CONVERTER: 16 bit resolution

10. DISCRETE OUTPUTS:

GMP1: Outputs 1 and 2 available as Solid State NFET or Form A relay.

Output 3 is an SSR NFET or Form C relay.

GMP2: Outputs 1 through 4 available as Form A relay or Solid State NFET.

Solid State Output:

Type: Switched DC, N Channel open drain MOSFET

Current Rating: 1 A max

VDS ON: 0.3 V @ 1 A

VDS MAX: 30 VDC

Offstate Leakage Current: 0.5 mA max

Form A Relay Output:

Type: N.O.

Current Rating: 3 Amps @ 125 VAC or 30 VDC (resistive load)

1/10 HP @ 125 VAC

Life Expectancy: 200,000 cycles at maximum load rating. (Decreasing

load, increasing cycle time, and use of surge suppression such as RC

snubbers increases life expectancy.)

Form C Relay Output:

Type: SPDT

Current Rating: 5 Amps @ 125 VAC or 28 VDC (resistive load)

1/8 HP @ 125 VAC

Life Expectancy: 100,000 cycles at maximum load rating. (Decreasing

load, increasing cycle time, and use of surge suppression such as RC

snubbers increases life expectancy.)

11. CONTROL MODES:

Control: On/Off, P, PI, or PID

Output: Time proportioning or linear (GMP1 only)

Cycle Time: Programmable from 0.0 to 60.0 sec

Auto-Tune: When selected, sets proportional band, integral time, derivative

time values, and output dampening time

Probe Break Action: Programmable response

Sensor Fail Response: Upscale

12. ALARMS:

Modes:

Manual

Absolute High Acting

Absolute Low Acting

Deviation High Acting

Deviation Low Acting

Inside Band Acting

Outside Band Acting

Reset Action: Programmable; automatic or latched

Standby Mode: Programmable; enable or disable

Hysteresis: Programmable

Sensor Fail Response: Upscale

13. ANALOG DC OUTPUT (optional, GMP1 only):

Selectable/programmable for 0-10 VDC, 0-20 mA, or 4-20 mA

Resolution:

Voltage: 500 µV

Current: 1 µA

Accuracy:

0.1% of full scale (18 to 28 °C)

0.2% of full scale (-40 to 75 °C)

Update Time: 0.0 to 60.0 sec

Compliance (for current output only): 500 Ω max.

Minimum load (voltage output only): 10 KΩ min.

Outputs are software selectable for either 10 V or 20 mA. The output range

may be field calibrated to yield approximate 10% overrange and a small

underrange (negative) signal.