TECHNICAL REFERENCE___________________________________________________________
34 __________________________________________________________________ M211097EN-C
Table 10
Required Parameters for WS425 Analog Output
Operation Mode
Parameter Name Default
Value
Allowed Values
Description
aoErrValue
1000
0 …. 32000
Set 2 for voltage output and 1000 for
frequency output for similar operation as
with WS425.
aout1_g
1
0 …. 100
Gain for AOUT1.
Set 0.01789 for compatibility with WS425
voltage output.
Set 11.18 for compatibility with WS425
frequency output.
aout2_g
1
0 …. 100
Gain for AOUT2.
Set 0.0027 for compatibility with WS425
potentiometer output.
aout1mode 3
(0
=
current)
1 = voltage
2 = frequency
(3 = disabled)
Analog output mode for AOUT1.
Set 1 for voltage and 2 for frequency.
aout2mode 7
(4
=
current)
(5 = voltage)
6 = potentiometer
(7 = disabled)
Analog output mode for AOUT2.
Set 6 for potentiometer.
aout1_o
aout2_o
0
-10000 … 10000
Offset for AOUT1 and AOUT2.
Set output offset to 0.
For configuration instructions, see Vaisala WINDCAP
®
Ultrasonic Wind
Sensor Series WMT700 User's Guide.