Figure 24, Dpt146 dewpoint measurement accuracy, Table 16 – Vaisala DPT146 User Manual

Page 64: Operating environment, Table 17, Outputs, Table 17 outputs

background image

User's Guide _______________________________________________________________________

62 ___________________________________________________________________ M211372EN-D

1109-009

Figure 24

DPT146 Dewpoint Measurement Accuracy

Table 16

Operating Environment

Property

Description / Value

Operating temperature of electronics

-40 … +60 °C (-40 ... +140 °F)

Pressure

0 … 50 bar (0…725 psi)

Relative humidity

0 ... 100 %

Measured gases

Air/non-corrosive gases

Sample flow rate

No effect on measurement accuracy

Table 17

Outputs

Property

Description / Value

Analog outputs (2 channels)

current output

0 … 20 mA, 4 … 20 mA

voltage output

0 … 5V, 0 …10V

Accuracy of analog outputs

± 0.01 V / ± 0.01 mA

Digital output

RS-485, non-isolated, Vaisala protocol

Connector

4-pin M8

0

20

20

0

10

30

-10
-20
-30

10

30 40 50 60

40

D

ew

po

in

t

te

m

pe

ra

tu

re

C

)

Temperature of measured gas (°C)

-40 -30 -20 -10

70 80

-40
-50
-60

Not recommended for

continuous measurement

Dewpoint measurement low limit

-70

Accuracy not specified

Accuracy

±2 °C Td

-80