DSC74C Two-Pressure Sampling Cell

with Coil

DSC74C with an outlet coil is designed for the most critical

measurements at atmospheric pressure. In the DSC74C, the coil is

connected to the sampling cell outlet to protect the sensor from ambient

humidity disturbing the measurement. The coil can also be used as a

cooling coil for sampling from high temperature processes.

DSC74C parts are:

- Sampling cell, thread G3/8"
- Connection part with a needle valve and an integrated leak screw
- Reducing Nipple (thread adapter), G3/8" - G1/2"
- Reducing Adapter (thread adapter), G3/8" - G1/4"
- Diffusion coil (for measurements in atmospheric pressure)

0510-034

Figure 17

Default Assembly of DSC74C

where
1 = Gas goes in. The coil can also be used here.

2 = Gas comes out

3 = Coil

4 = Valve