Chapter 4 ________________________________________________________________ Installation

VAISALA ________________________________________________________________________ 31

DM240FA Duct Installation Flange

The transmitter can be installed directly in the measured gas through

DM240FA duct installation flange. When the transmitter is installed

directly on the process wall or pipe, note that a closing valve may be

needed on both sides of the installed transmitter so that the transmitter

can be removed from the process for calibration or maintenance.

1108-006

Figure 19

DM240FA with DPT146

where
1 = Measured gas

2 = DPT146 transmitter

3 = DM240FA flange (thread G1/2" ISO)

4 = Recommended additional hole (plugged) for dewpoint field

check reference measurement probe (for example, Vaisala

DMP74B)