Set Analog Output Parameters and

Scaling

Use the ASEL command to show or set the output parameters and the

scalings of the analog output.

ASEL [ch1] [ch2]

where
ch1 = Parameter that is output on analog channel 1. Available

parameters are TDF, TDFA, H2O, and P.

ch2 = Parameter that is output on analog channel 2. Available

parameters the same as for channel 1.

Example (show current settings):

asel

Ch1 Tdf lo : -60.00 'C ?

Ch1 Tdf hi : 40.00 'C ?

Ch2 P lo : 0.00 bara ?

Ch2 P hi : 10.00 bara ?

Example (change output parameters to TDFA and H2O and set scaling):

asel h20 ta

Ch1 Tdfa lo : 00.00 'C ? -80

Ch1 Tdfa hi : 00.00 'C ? 20

Ch2 H2O lo : 0.00 ppm ? 0

Ch2 H2O hi : 0.00 ppm ? 500

Set Analog Output Error Notification

If the transmitter is malfunctioning, the analog output is set to a specified

level. This overrides the normal measurement output of the channel. The

default level is 0 mA or 0V, according to the output type. You can set the

level using the AERR command.

AERR [ch1] [ch2]

where
ch1 = Error level for analog channel 1.

ch2 = Error level for analog channel 2.

Example (show current settings):

aerr

Ch1 error out : 0.000 mA ?

Ch2 error out : 0.000 V ?