User's Guide _______________________________________________________________________

22 ___________________________________________________________________ M211372EN-D

measurement point should be within 1 ... 10 bar (14.5 ... 145 psi). The

maximum pressure tolerance of the transmitter is 50 bar (725 psi).

If the gas temperature is higher than the specified maximum operating

temperature of the transmitter, gas sampling and cooling it to ambient

temperature (for example, 20 °C / 68 °F) is recommended. Note that the

dewpoint temperature must be clearly lower than the ambient

temperature to avoid condensation in the sampling line. Sampling from

the process is easy by using Vaisala sampling cell options; see section

Sampling Accessories on page 25.

DPT146 is light in weight, which means that it can be installed in a

sample pipeline in the sampling cells without the need of any additional

mechanical support.

Installing the Transmitter

After selecting a suitable measurement location, follow the procedure

below to install the transmitter:

1.

Remove the yellow transport protection cap from the transmitter.

1109-030

Figure 8

Removing the Transport Protection Cap

2.

Make sure that the threads on the mounting point are of the correct

type, and fasten the transmitter to the measurement point. When

installing a transmitter with ISO or UNF thread, use the provided

sealing ring to achieve a tight installation. Check that the sealing

ring remains centered.