Chapter 5 _________________________________________________________________ Operation

VAISALA ________________________________________________________________________ 43

Outputting the results continues in intervals issued with the command

INTV. You can stop the output by pressing the Esc key, or with the

command S. Since the interface is half-duplex, you must enter the

commands when the transmitter is not outputting.

Stop Measurement Output

You can stop the measurement output by pressing ESC, or with the S

command:

S

Set Output Interval

Use the INTV command to change the output interval of the

automatically repeating measurement messages. The measurement

messages are repeated in the RUN mode, or after the R command has

been given.

INTV [n xxx]

where
n

= time interval, range 0 ... 255

xxx

= time unit = "S", "MIN", or "H"

The shortest output interval (with n = 0) outputs the measurement

messages as quickly as the transmitter’s internal measurement cycle

produces them, without additional delay.

Example:

intv 1 min

Output interval: 1 MIN