Chapter 16 _____________________________________________________ Serial Line Commands

VAISALA _______________________________________________________________________ 93

SCAL Starts the Salt Detection Limit Determination
Cycle

Syntax:

SCAL

The determination of the salt detection limit can take several minutes.
Please also see page 65 for more information about salt detecting.

Example:

>scal
Salt limit : 0.36280
>

PTEST Starts the Cooling Capacity Test

Syntax:

PTEST

During the test device reports the sensor temperature. When the test is
completed, the lowest achievable stable sensor temperature (in current
ambient conditions) is reported.

Example:

>ptest
-12.34 'C
-22.45 'C
-31.32 'C
-35.44 'C
...
-39.34 'C
Cool. capacity: -39.12 'C
>

Typing s
interrupts the test.