User's Guide ______________________________________________________________________
62 __________________________________________________________________ M210205EN-D
Response Speed
You can set the response speed to stable, normal (default) or fast. If
fast transients in the dewpoint prevent the system from reaching
balance, the signal can be stabilized by selecting the Stable response
speed. Fastest response time is achieved by setting the speed to Fast
but this may increase the noise in the dewpoint measurement. As a
rule of thumb, always use the fastest setting that gives stable readings!
1.
Open
MENU
: press .
2.
Select ►
Measuring
, press .
3.
Select ►
Measurement settings
, press .
4.
Select
Response
, press
SET
.
5.
Select one of the following options:
Stable
: Press
SELECT
to have the slowest response speed.
Normal
: Press
SELECT
to have normal response speed.
Fast:
Press
SELECT
to have fast response speed.
6.
Press
EXIT
to return to the main view.
Pump ON/OFF
1.
Open
MENU
: press .
2.
Select ►
Measuring
, press .
3.
Select ►
Measurement settings
, press .
4.
Select
Pump
, press
ON/OFF
.
5.
Press
EXIT
to return to the main view.