Vaisala DM500 User Manual
Page 94
User's Guide ______________________________________________________________________
92 __________________________________________________________________ M210205EN-D
HEAT Starting/Stopping the Optional Heater
Syntax:
HEAT ON/OFF
Example:
>heat
Heat : OFF
>
>heat on
Heat : ON
>
>heat off
Heat : OFF
>
SALT Enables/Disables the Salt Contamination
Detection
Syntax:
SALT ON/OFF [interval]
Default: SALT ON
Example:
>salt
Salt sensing : OFF 24
>
Salt sensing is off.
>salt on
Salt sensing : ON 24
>
Salt sensing is on and detection interval is 24 h.
>salt 30
Salt sensing : ON 30
>
Salt sensing is on and detection interval is 30 h.
>salt on 24
Salt sensing : ON 24
>
Salt sensing is on and detection interval is 24 h.
See also other documents in the category Vaisala Tools:
- DM70 (93 pages)
- DMT132 (74 pages)
- DMT143 (76 pages)
- DMT152 (70 pages)
- DMT242 (4 pages)
- DMT340 (191 pages)
- DMT345 (185 pages)
- DPT145 (63 pages)
- DPT146 (71 pages)
- PTU300 (217 pages)
- PTB330TS (89 pages)
- PTB220 (10 pages)
- PTB220 (113 pages)
- PTB330 (144 pages)
- PTU200 (64 pages)
- PTU200MIK1 (18 pages)
- SPH10 (2 pages)
- SPH20 (2 pages)
- PTB110 (4 pages)
- PTB200 (30 pages)
- PTB210 (analog) (27 pages)
- PTB210 (serial) (32 pages)
- GM70 (68 pages)
- GMD20 (4 pages)
- GMK220 (18 pages)
- GML20 (2 pages)
- GML20T (2 pages)
- GMM20W (5 pages)
- GMM220 (6 pages)
- GMP231 (90 pages)
- GMP231 (2 pages)
- GMP343 (94 pages)
- GMR20 (2 pages)
- GMT220 (42 pages)
- GMW90 (101 pages)
- XMW90 (4 pages)
- MM70 (67 pages)
- MM70 (71 pages)
- MMT162 (66 pages)
- MMT310 (81 pages)
- MMT330 (181 pages)
- MMT330 (171 pages)