Vaisala DM500 User Manual
Page 105

Chapter 16 _____________________________________________________ Serial Line Commands
VAISALA ______________________________________________________________________ 103
The CDP* command displays also the coefficients of the correction
polynomial, as well as the reference value and residual error for each
calibration data point.
Example:
>cdp
1. reference : -39.99
1. measurement : -40.009
2. reference : -30.01
2. measurement : -30.030
3. reference : -20.01
3. measurement : -20.026
4. reference : -9.98
4. measurement : -9.977
5. reference : 0
5. measurement : -0.006
6. reference : 10
6. measurement : 10.002
7. reference : 19.99
7. measurement : 19.982
8. reference : 29.98
8. measurement : 29.989
9. reference : 39.98
9. measurement : 39.989
10. reference :
Reference Measurement
-39.990 -40.009
-30.010 -30.030
-20.010 -20.026
-9.980 -9.977
0.000 -0.006
10.000 10.002
19.990 19.982
29.980 29.989
39.980 39.989
Overwrite previous calibration ? Y/N
OK
>
>cdp *
1. reference : -39.99
1. measurement : -40.009
2. reference : -30.01
2. measurement : -30.030
3. reference : -20.01
3. measurement : -20.026
4. reference : -9.98
4. measurement : -9.977
5. reference : 0
5. measurement : -0.006
6. reference : 10
6. measurement : 10.002
7. reference : 19.99