Operational commands – Vaisala DM500 User Manual
Page 93
![background image](/manuals/408481/93/background.png)
Chapter 16 _____________________________________________________ Serial Line Commands
VAISALA _______________________________________________________________________ 91
3.
Output needed: Relative humidity and pressure to be shown with
a unit. The fields are separated with " " marks. Carriage return
and line feed in the end.
>form 2.1 rh " " u " " 2.4 p " " u \r \n
2.1 RH " " U " " 2.4 P " " U \r \n
>send
23.0 % 1.0164 bar
Operational Commands
START Starting the Measurement
Syntax:
START
Starts the dewpoint measurement and activates the cooling system.
STOP Stopping the Measurement
Syntax:
STOP
Stops the dewpoint measurement and disables the cooling system.
PUMP Starting/Stopping the Optional Sample Gas
Pump
Syntax:
PUMP ON/OFF
Example:
>pump
Pump : OFF
>
>pump on
Pump : ON
>
>pump off
Pump : OFF
>
- DM70 (93 pages)
- DMT132 (74 pages)
- DMT143 (76 pages)
- DMT152 (70 pages)
- DMT242 (4 pages)
- DMT340 (191 pages)
- DMT345 (185 pages)
- DPT145 (63 pages)
- DPT146 (71 pages)
- PTU300 (217 pages)
- PTB330TS (89 pages)
- PTB220 (10 pages)
- PTB220 (113 pages)
- PTB330 (144 pages)
- PTU200 (64 pages)
- PTU200MIK1 (18 pages)
- SPH10 (2 pages)
- SPH20 (2 pages)
- PTB110 (4 pages)
- PTB200 (30 pages)
- PTB210 (analog) (27 pages)
- PTB210 (serial) (32 pages)
- GM70 (68 pages)
- GMD20 (4 pages)
- GMK220 (18 pages)
- GML20 (2 pages)
- GML20T (2 pages)
- GMM20W (5 pages)
- GMM220 (6 pages)
- GMP231 (2 pages)
- GMP231 (90 pages)
- GMP343 (94 pages)
- GMR20 (2 pages)
- GMT220 (42 pages)
- GMW90 (101 pages)
- XMW90 (4 pages)
- MM70 (67 pages)
- MM70 (71 pages)
- MMT162 (66 pages)
- MMT310 (81 pages)
- MMT330 (171 pages)
- MMT330 (181 pages)