beautypg.com

Technical data, Specifications, Chapter 6 – Vaisala HMDW80 User Manual

Page 47: Table 9, Specifications for hmw82/83 and tmw82/83, Chapter 6, technical data, Ee chapter 6, technical, Data, on pa

background image

Chapter 6 _____________________________________________________________ Technical Data

VAISALA ________________________________________________________________________ 45

CHAPTER 6

TECHNICAL DATA

This chapter provides the technical data of the HMDW80 series.

Specifications

Table 9

Specifications for HMW82/83 and TMW82/83

Property

Description / Value

Operating environment

Operating temperature range

-5 ...+55 °C (+23 ... +131 °F)

Operating humidity range

0 …100 %RH, non-condensing

Relative humidity

Measurement range

0 … 100 %RH

Accuracy

Temperature range

+10 ... +30 °C (+50 ... +86 °F)

30 ... 70 %RH

±3 %RH

0 … 30 %RH, 70 … 100 %RH

±5 %RH

Temperature range

-5 ... +10 °C, +30 ... + 55 °C

(+23 ... +50 °F, +86 ... +131 °F)

0 …100 %RH

±7 %RH

Stability in typical HVAC applications

±2 %RH over 2 years

Humidity sensor

Vaisala INTERCAP

®

Temperature

Measurement range

-5 ...+55 °C (+23 ... +131 °F)

Accuracy

+10 … +30 °C (+50 °F … +86 °F)

±0.5 °C (±0.9 °F)

-5 … +10 °C, +30 … +55 °C

(+23 ... +50 °F, +86 ... +131°F)

±1.0 °C (±1.8 °C)

Temperature sensor

Digital temperature sensor