Replacing the intercap, Sensor on wall models, Hmw82/83 – Vaisala HMDW80 User Manual
Page 37: Figure 21, Intercap, Sensor on hmw82/83 models, Sensor on wall models hmw82/83

Chapter 4 _______________________________________________________________ Maintenance
VAISALA ________________________________________________________________________ 35
Replacing the INTERCAP
®
Sensor on
Wall Models HMW82/83
REQUIRED
TOOLS
- New INTERCAP
®
humidity sensor.
- Small flat head screwdriver.
1.
Use the screwdriver to push down on the tab on the top of the
transmitter and open the transmitter cover.
2.
Remove the old sensor, and insert the new one. See Figure 21
3.
Reconnect the cover to the mounting base.
4.
Verify that the transmitter now provides a valid humidity reading.
1305-018
Figure 21
INTERCAP
®
Sensor on HMW82/83 Models
1
See also other documents in the category Vaisala Humidifiers:
- Calibration of Digital Transmitters with HMI41 (36 pages)
- Calibration of Series HMDW2030 and HMP130 Transmitter with HMI41 (14 pages)
- Calibration of Series HMDW6070 and HMP140 Transmitter with HMI41 (30 pages)
- HM34 (30 pages)
- HM40 (47 pages)
- HM44 (52 pages)
- HM70 (83 pages)
- HMD40 (1 page)
- HMD60 (4 pages)
- HMD70 (18 pages)
- HMDW110 (62 pages)
- HMI41 (74 pages)
- HMP41 (72 pages)
- HMK15 (39 pages)
- HMM100 (71 pages)
- HMM105 (23 pages)
- HMM211 (42 pages)
- HMM212 (36 pages)
- HMM213 (52 pages)
- HMP140 (28 pages)
- HMP155 (84 pages)
- HMP228 (115 pages)
- HMP230 (163 pages)
- HMP240 (130 pages)
- HMP260 (118 pages)
- HMP60 (71 pages)
- HMT100 (52 pages)
- HMT120 (87 pages)
- HMT130 (95 pages)
- HMT140 (76 pages)
- HMT310 (88 pages)
- HMT310 (105 pages)
- HMT330 (209 pages)
- HMT360 (97 pages)
- HMT360 (63 pages)
- HMT360N (110 pages)
- HMW40 (1 page)
- HMW90 (110 pages)
- SHM40 (68 pages)
- RDP100 (14 pages)