beautypg.com

Analog output error state, Error messages on the display, Table 8 – Vaisala HMDW80 User Manual

Page 44

background image

User's Guide _______________________________________________________________________

42 ___________________________________________________________________ M211510EN-C

Analog Output Error State

HMDW80 series transmitters set the output channel(s) into a defined

error level instead of the measured result in two situations:

- Transmitter detects a measurement malfunction. This means an actual

measurement problem, such as sensor damage or unsuitable

environmental conditions.

- Measured value(s) are well outside the scaled output range. For a

more detailed explanation, see section Analog Output Overrange

Behavior on page 16.

The error level depends on the output type:
- For 0 ... 10 V output, the error level is 11 V.

HMW83 and TMW83 are an exception to this, their error level is 0 V.

- For 4 ... 20 mA output, the error level is 3.6 mA.

The transmitter resumes normal operation of the analog output(s) when

the cause of the error state is removed.

Error Messages on the Display

Table 8

Error Messages on the Display

Error Message

Possible Cause and Solution

INTERNAL 1

INTERNAL 2

INTERNAL 3

Internal problem with the transmitter. Restart the

transmitter to see if the problem remains. Contact Vaisala

technical support if unable to remove the problem.

ERROR HUM

Humidity measurement failed. Check if the humidity

sensor is damaged or missing.

ERROR T

Temperature measurement failed. Check if the

temperature sensor is damaged or missing.