Chapter 4 _______________________________________________________________ Maintenance

VAISALA ________________________________________________________________________ 33

CHAPTER 4

MAINTENANCE

This chapter provides information that is needed in basic maintenance of

HMDW80 series.

Cleaning

The body of the transmitter can be cleaned by wiping with a moistened

lint-free cloth. Do not use cleaning agents or solvents, or blow

pressurized air into the transmitter housing or on the filter.

NOTE

Do not attempt to clean contaminated INTERCAP

®

sensors or filters.

Instead, replace them with new parts. Filters and sensors can be

purchased from Vaisala. For order codes, see section Spare Parts and

Accessories on page 48.

Calibration

You can verify the performance of your HMDW80 series transmitter by

comparing its reading with the reading from a portable reference

instrument such as the HM70 Hand-Held Humidity and Temperature

Meter. For operating instructions, refer to the documentation of your

reference instrument.

If the comparison indicates that the reading from the transmitter is not

within specification, the transmitter may need to have its filter and sensor

replaced.