beautypg.com

Installing wall models hmw88 and hmw89, Figure 11, Wall model dimensions (hmw88/89) – Vaisala HMDW80 User Manual

Page 22: Required tools

background image

User's Guide _______________________________________________________________________

20 ___________________________________________________________________ M211510EN-C

Installing Wall Models HMW88 and HMW89

REQUIRED

TOOLS

- Medium size crosshead screwdriver (Pozidriv) for cover screws.
- Small slotted screwdriver for screw terminals.
- Two installation screws: Ø < 3.5 mm, head Ø < 10 mm.
- Depending on the wall material and screw type, you may need a

drill and a suitable drill bit to make installation holes for screws.

- Tools for cutting and stripping wires.
- 19 mm open-end wrench for tightening the cable gland.


1.

Open the transmitter cover and use two screws (not included) to

attach the transmitter to the wall. The probe and cable gland should

point down.

2.

Route the power and signal cable to the screw terminals and

connect the wires. Refer to section Wiring on page 26.

3.

Close the transmitter cover and remove the yellow transport

protection cap from the probe.

1302-029

Figure 11

Wall Model Dimensions (HMW88/89)

88

Ш3.5

Ш10

Ш3.5

Ш10

192.

5

72

40.6

Ш4...8

105

87.

5

Ш12