Troubleshooting, Problem situations, Chapter 5 – Vaisala HMDW80 User Manual

Page 43: Table 7, Possible problem situations and their remedies, Chapter 5, troubleshooting

background image

Chapter 5 ____________________________________________________________ Troubleshooting

VAISALA ________________________________________________________________________ 41

CHAPTER 5

TROUBLESHOOTING

This chapter describes possible problems, their probable causes and

remedies, and provides contact information for technical support.

Problem Situations

Table 7

Possible Problem Situations and Their Remedies

Problem

Probable Cause and Remedy

Transmitter outputs an

incorrect humidity reading.

1. Models with probes: check if the yellow

transport protection cap has been left over

the probe. Remove it if present.

2. Models with configurable output parameter:

check that the desired output parameter

has been selected using the DIP switches.

Temperature reading shown

by the transmitter is too high.

1. The transmitter may be installed in an

unsuitable location, for example, near a

heat source or in sunlight. Use a portable

reference meter to check this.

2. HMW82 and HMW83: Check that the

transmitter is installed in proper orientation,

with the arrow on the mounting base

pointing up.

Analog output reading is

unchanging and incorrect.

The analog output channel may be in an

error state. See section Analog Output Error

State on page 42.