Introduction – Posey Bed 8070 User Manual

Page 4

background image

4

Posey Bed - 8070

Introduction

Power Supply Requirements

Model:

Posey Bed 8070

Rated voltage

and current:

120 VAC

Frequency:

50-60Hz

Current Rating:

3.5 Amps Max

Classification:

Class II, Type B (Double

insulated)

Mode of Operation:

10% Maximum duty cycle

Dimensions

Highest Position

Inches:

84.25(W) x 40.875(D) x 73.25(H)

Centimeters:

214(W) x 104(D) x 186(H)

Lowest Position

Inches:

84.25(W) x 40.875(D) x 62.75(H)

Centimeters:

214(W) x 104(D) x 159(H)

Area:

19.5 Square Feet

Sleep Surface
Volume Inside:

70 Cubic Feet (2 cubic meters)

Maximum Back Angle: 65° to Horizontal
Maximum Knee Angle: 35° to Horizontal
Maximum Knee

to Leg Angle:

45° to Horizontal

Unit Weight
Canopy and Frame:

86 lbs. (39 kgs.)

Mattress:

20 lbs. (9 kgs.)

Invacare Platform:

150 lbs. (68 kgs.)

Assembled Bed:

256 lbs. (116 kgs.)

Operating Conditions

The Posey Bed is designed for use in normal indoor

environments.

Storage Conditions

The Posey Bed may be stored in ambient warehouse

temperatures at normal humidity levels. Avoid excess

moisture or high humidity that may damage product

materials.

Safe Working Load

Weight limitations:

*450 lbs. (205 kgs.)

*includes all accessories, mattress and canopy.

Regulatory Listing

UL 60601-1

IEC 60601-2-38

IEC 60601-1-2

Specifications for Invacare SC900DLX Bed modified for the Posey

Bed

• “MEDICAL ELECTRICAL EQUIPMENT needs

special precautions regarding EMC and needs to

be installed and put into service according to

the EMC information provided in this manual.”

• “Portable and mobile RF communications

equipment can affect MEDICAL ELECTRICAL

EQUIPMENT.”

• “The use of accessories, transducers and

cables other than those specified by Posey

Company, may result in increased EMISSIONS or

decreased IMMUNITY of the Posey Bed 8070.”

• “This Posey Bed 8070 should not be used

adjacent to or stacked with other equipment

and that if adjacent or stacked use is necessary,

the Posey Bed 8070 should be observed to

verify normal operation in the configuration in

which it will be used.”