Posey Exit Alarm Mat User Manual

Posey, Exit mat sensor, Before you begin

background image

To reduce the risk of serious injury or death:
• Read Sensor and Alarm instructions prior to use.

ALWAYS test sensor before leaving patient unattended.

– Apply pressure to pad and then release. Do this at

several places along the entire length of the mat.

Alarm should activate each time you apply pressure.
– If sensor fails any test, stop use at once and replace

with a new sensor.
– Sensors past warranty expiration date may fail, causing

false alarm or no alarm.

NEVER jerk or pull on cable to remove RJ11 plug. Doing

so will damage the cord or plug and may cause sensor to

fail.

ALWAYS use plastic tab to release plug.

NEVER roll, fold or crease

sensor. This will damage

sensor and may cause false

alarm or no alarm.

• This device should not be a

substitute for maintaining the routine visual monitoring

protocol by healthcare personnel.

• Test before placing this sensor under any other floor

mat, as it may not work. DO NOT leave patient

unattended if floor mat sensor does not activate alarm.

8250L

Exit Mat Sensor

Designed to work with Posey Fall Monitors, the Posey Exit

Mat Sensor provides caregivers an early warning of patient

attempts to exit a bed, through a doorway, or enter

“off-limit” areas. The Exit Mat Sensor is available in two

sizes for use either next to a patient’s bed, or in front of a

door.

REF

8250S Exit Mat Sensor, Small - 36”L x 24”W (91 cm x 61 cm)

REF

8250L Exit Mat Sensor, Large - 60”L x 24”W (152 cm x 61 cm)

Before You Begin

When used properly, Posey sensors and fall alarms alert

staff that a patient has applied weight to the sensor and

may be at immediate risk of falling. However, fall alarm

systems DO NOT prevent falls and are not a substitute for

good nursing care and regular visual monitoring.

• This device may not be suitable for all high fall-risk

patients. See Posey catalog for other options for such

patients.

NEVER connect Posey sensors to other manufacturers’

alarm units.

NEVER connect Posey Alarms to other manufacturers’

sensors.

Application Instructions:

1. Position the Exit Mat Sensor on the floor, next to the

bed, or in front of a doorway.

2. Connect the sensor to a Posey Fall Monitor.
3. Turn Fall Monitor ON.
4. Mount the Fall Monitor on the headboard or wall using

the brackets included with the Fall Monitor.

5. Tuck excess sensor cable under the corner of the mat or

out of the way to avoid a tripping hazard.

System Test:

1. Plug sensor into Alarm unit.
2. Turn Alarm unit on and verify power light is lit.
3. Check that audio alarm is functioning properly by

stepping on the Sensor and activating the Alarm.

4. Ensure the nurse call interface, when used, is connected

and activates properly.

5. Please refer to the Alarm unit owner’s manual for

additional information.

© 2010 Posey Company. All rights reserved.

M1409 REV D 090110

Posey Company • 5635 Peck Road, Arcadia, CA 91006-0020 USA • www.posey.com

Phone: 1.800.447.6739 or 1.626.443.3143 • Fax: 1.800.767.3933 or 1.626.443.5014

Posey

®

Exit Mat Sensor

FALL MANAgEMENt