Guidance and manufacturer’s declaration – Posey Bed 8070 User Manual

Page 24

background image

Table 202

Guidance and manufacturer’s declaration – electromagnetic immunity

The Posey Bed 8070 is intended for use in the electromagnetic environment specified below. The

customer or the user of the Posey Bed 8070 should assure that it is used in such an environment.

Immunity Test

IEC 60601

Compliance Level

Electromagnetic Environment –

Test Level

Guidance

NOTE UT is the a.c. mains voltage prior to application of the test level.

Electrostatic

discharge (ESD)

IEC 61000-4-2

Electrical fast

transient/burst

IEC 61000-4-4

Surge

IEC 61000-4-5

Voltage dips, short

interruptions and

voltage variations

on power supply

input lines

IEC 61000-4-11

(50/60 Hz)

magnetic field

IEC 61000-4-8

+

-6 kV contact

++

-8 kV air

+

-2 kV for power

supply lines

+

-1 kV for input/

output lines

+

-1 kV differential

mode

+

-2 kV common

mode
<5% UT

(>95% dip in UT)

for 0,5 cycle

40% UT

(60% dip in UT)

for 5 cycles

70% UT

(30% dip in UT)

for 25 cycles

<5% UT

(>95% dip in UT)

for 5 sec

3 A/m

+

-6 kV contact

+

-8 kV air

+

-2 kV for power

supply lines

+

-1 kV for input/output

lines

+

-1 kV differential

mode

+

-2 kV common mode

<5% UT

(>95% dip in UT)

for 0,5 cycle

40% UT

(60% dip in UT)

for 5 cycles

70% UT

(30% dip in UT)

for 25 cycles

<5% UT

(>95% dip in UT)

for 5 sec

3 A/m

Floors should be wood, concrete or

ceramic tile. If floors are covered with

synthetic material, the relative

humidity should be at least 30%.
Mains power quality should be that of

a typical commercial or hospital

environment.

Mains power quality should be that of

a typical commercial or hospital

environment.

Mains power quality should be that of

a typical commercial or hospital

environment. If the user of the

Posey Bed 8070 requires

continued operation during power

mains interruptions, it is

recommended that the Posey

Bed 8070 be powered from an

uninterruptible power

supply or a battery.

Power frequency magnetic fields

should be at levels characteristic of a

typical location in a typical

commercial or hospital environment.

24

Posey Bed - 8070

Guidance and Manufacturer’s Declaration