Guidance and manufacturer’s declaration – Posey Bed 8070 User Manual
Page 26
Table 206
Recommended separation distances between portable and mobile RF
communications equipment and the Posey Bed 8070
The Posey Bed 8070 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Posey Bed 8070 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Posey Bed 8070 as recommended below, according to
the maximum output power of the communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Separation distance according to frequency of transmitter
m
Rated maximum
output power
of transmitter
W
0.01
0.1
1
10
100
150 kHz to 80MHz
d = [3.5/V1]√P
0.12
0.37
1.17
3.69
11.67
80 MHz to 800 MHz
d = [3.5/E1]√P
0.12
0.37
1.17
3.69
11.67
800 MHz to 2.5 GHz
d = [7/E1]√P
0.23
0.74
2.33
7.39
23.33