2 system management, 1 self check, System management – Boonton PIM 31 User Manual User Manual
Page 65: Self check
![background image](/manuals/401811/65/background.png)
65
8.2 System Management
8.2.1 Self Check
Self check provides information about the PIM 31 test system. To
allow the self check routine to test all modules, it needs to be
performed with RF Power ON
Name
Description
HPAx Output
RF Output power level (should be +/- 0.2
db within the set limits – ALC ON)
HPAx ON/OFF
Status of RF Power (ON/OFF)
HPAx Temperature
Temperature of amplifiers.
HPAx Over Power
Reverse Power
HPAx Temp Status
On Fail (too hot), System will shut down
Synthesizerx Lock
Synthesizers of Transmitter and Receiver
locked
HPAx Reverse
Reverse PIM measurement blocks status
Local Osc Lock
Local Oscillator functionality
Fan Operating
Operation status of 6 fans