Table 3-8. Sensor Zero / Cal Status Codes
00
operation complete, no errors
03
coarse peak offset failed
04
top power code too low or too high
05
zero power point invalid
06
power entries nonmonotonic
07
code entries nonmonotonic (sensor may not be warmed up - try again)
08
too far (over 32dB) to extend table
13
zero offset signal level too low
14
zero offset signal level too high (sensor probably connected to signal source)
15
zero offset code value out of range (sensor probably connected to signal source)
23
step calibration signal level too high
24
zero offset adjustment invalid
26
autocal step signal level too high
27
autocal step signal level too low (sensor probably not connected to correct calibrator)
28
step calibration linearity questionable (sensor may not be warmed up - try again)
29
fixed cal power level invalid (fixed cal should be attempted only at 0dBm)
2A
fixed cal input power level too high
2B
fixed cal input power level too low
Table 3-9. Startup Error Messages
These messages appear briefly during startup if a system error is detected. The error status may also be viewed on
the display menu Utilities>Sys-Tests>SystemInf. Try a power off/on cycle to recover normal operation. If this fails,
service may be required.
DSP S/W Failed
The DSP is not running or failed to respond.
DSP - NoResp
During normal operation, the DSP failed to respond to a command.
DSP-CalTbl
During the boot-up process, an error occurred while down loading the calibra-
tion tables to the DSP.
DSP-IniErr
During the boot-up process, an error occurred while down loading the channel
parameters to the DSP.
No System Calib. Tables
The calibrator is missing its internal calibration tables. This requires factory
recalibration of the internal 50 MHz calibrator for proper operation.
No DC Calibration Data
The instrument is missing its CW channel gain calibration table. This requires
factory recalibration for proper operation.
Chan# Cal Table: Err
The DSP cal table for this peak sensor is not initialized. An autocal is needed.
Sensor# EEProm: Err
Checksum Error in one of the EEPROM pages. If this error occurs again after a
power cycle, sensor service may be required.
No DSP Data Records
The flash memory does not contain valid DSP instruction code. Try reinstalling
operating firmware.