SENSe[1 | 2]:PULse:DISTal

set/return pulse distal level in percent of top power, = 0 to 100%

SENSe[1 | 2]:PULse:MESIal

set/return pulse mesial level in percent of top power, = 0 to 100%

SENSe[1 | 2]:PULse:PROXimal

set/return pulse proximal level in % of top power, = 0 to 100%

SENSe[1 | 2]:PULse:UNITs

set/return power units used for DIST, MESI and PROX arguments

SENSe[1 | 2]:PULse:STARTGT

set/return pulse gate start in % of pulse duration, = 0 to 40%

SENSe[1 | 2]:PULse:ENDGT

set/return pulse gate end in % of pulse duration, = 60 to 100%

SENSe[1 | 2]:HIST:COUNt

set/return the number of histogram bins to return, = 0 to 4096

SENSe[1 | 2]:HIST:INDEX

set/return index of first histogram bin to return, = 0 to 4095

SENSe[1 | 2]:HIST:DATA?

return “count” bin values from the signal histogram, starting at “index”

SENSe[1 | 2]:CALTAB:COUNt

set/return the number of cal table values to return, = 0 to 4096

SENSe[1 | 2]:CALTAB:INDEX

set/return index of first cal table value to return, = 0 to 4095

SENSe[1 | 2]:CALTAB:DATA?

return “count” power values from the cal table, starting at “index”.

SENSe[1 | 2]:MBUF:SIZe

set/return the number readings needed to fill meas buffer, = 0 to 4096

SENSe[1 | 2]:MBUF:RATe

set/return meas buffer rate for Mod & CW, = 1 to 1000 rdgs/sec

SENSe[1 | 2]:MBUF:POSition?

return current acquisition index position of measurement buffer

SENSe[1 | 2]:MBUF:COUNt

set/return the number of meas buffer readings to return, = 0 to 4096

SENSe[1 | 2]:MBUF:INDEX

set/return index of first meas buffer reading to return, = 0 to 4095

SENSe[1 | 2]:MBUF:DATA?

return “count” stored readings from the meas buffer, starting at “index”

SENSe[1 | 2]:SBUF:MODE

set/return status of user defined sample buffer, = ON, OFF

SENSe[1 | 2]:SBUF:PERiod

set/return the A/D sampling period, = 5 to 12,500

SENSe[1 | 2]:SBUF:PREsamp

set/return the number of samples before the trigger, = 0 to 12,000

SENSe[1 | 2]:SBUF:POSTsamp

set/return the number of samples after the trigger, = 0 to 12,000

SENSe[1 | 2]:SBUF:COUNt

set/return the total number of samples to return, = 1 to 12,000

SENSe[1 | 2]:SBUF:INDEX

set/return the index of next data point relative to trigger position

SENSe[1 | 2]:SBUF:DATA?

return “count” power samples from captured buffer starting at “index”.

SENSe[1 | 2]:TEMPerature?

return sensor internal temperature in degrees Celsius

SENSe[1 | 2]:CALTemp?

return sensor calibration temperature in degrees Celsius

STATus:PRESet

set device dependent SCPI and device status registers

STATus:OPERation:EVENt?

return value of Operation Status Register

STATus:OPERation:CONDition?

return value of Operation Condition Register (unlatched)

STATus:OPERation:ENABle

set/return value of Operation Enable Mask, = 0 to 32767

STATus:QUEStionable:EVENt?

return value of Questionable Status Register

STATus:QUEStionable:CONDition?

return value of Questionable Condition Register (unlatched)

STATus:QUEStionable:ENABle

set/return value of Questionable Enable Mask, = 0 to 32767

SYSTem:BEEP

set/return keypad audible beeper status, = ON, OFF

SYSTem:ERRor[:NEXT]?

return next queued error code number and description

SYSTem:COMMunicate:SERial:BAUD

set/return serial port baud rate, = 1200 to 38400 baud

SYSTem:COMMunicate:SERial:BITS

set/return number of serial port data bits, = 7 or 8

SYSTem:COMMunicate:SERial:SBITs

set/return number of serial port stop bits, = 1 or 2

SYSTem:COMMunicate:SERial:PARity

set/return state of serial port parity bit, = NONE, ODD, EVEN

SYSTem:COMMunicate:SERial:CONTrol:RTS

set/return serial RTS handshake mode: OFF, ON