Boonton 4530 Peak Power Meter User Manual User Manual
Page 140

4-66
Chapter 4
Boonton Electronics
Remote Operation
4530 Series RF Power Meter
b. Making successive measurements.
C
ALC1:MODE cw
Set for CW measurement mode
TRIG:SOUR bus
Set trigger source for bus
ABORT
Measurements STOP and triggers go to idle.
{send GET} or *TRG
One random trace appears in graph mode
FETCH1:CW:POWER?
Returns CW power reading. Sending GET again has no effect;
the measurement is complete.
(The bus trigger serves no purpose unless an ABORT command is issued to reset the trigger system and stop the
measurement. Otherwise, just set
TRIGger:SOURce
immediate and use FETCH to input readings on demand. )
4.8.4 Statistical Mode Example
- (CDF, CCDF, DISTRIBUTION) - Signal in: CDMA signal at 0dBm average
power, at a 900MHz center frequency.
a. Start a CCDF measurement.
CALC1:MODE CCDF
Set CCDF display mode.
TRIG:SOUR immediate
Set for “immediate” start on trigger arming
ABORT
Reset the trigger system.
INIT:CONT off
Set for single, triggered measurement
INIT:IMMEDIATE
Restart statistical processing.
FETCH1:AMEASURE:POWER?
Return statistical reading array.
b. Using the bus trigger to start a CCDF measurement.
CALC1:MODE CCDF
Set CCDF display mode
TRIG:SOUR bus
Set trigger source for bus
ABORT
Reset trigger system.
INIT:CONT off
Set for single, triggered measurement
INIT:IMMEDIATE
Reset measurement.
{send GET or *TRG}
Start the measurement
FETCH1:AMEASURE:POWER?
Return statistical reading array