Problem, Possible cause, Solution – AirSep MN179-1 B User Manual

Page 56

background image

MN179-1 _ 02/14 AirSep Corp. - FreeStyle 5 Service Manual

Rev B

Problem

Possible Cause

Solution

Oxygen concentration
is below specifications
in test mode.

Unit’s temperature is too
high. Blocked air intake or
outlet; or defective fan.

Leak

Restriction in exhaust
felt.



Compressor has reached its

expected life.

Contaminated absorbent beds.

Defective solenoid valve.

Clear any obstruction in

air intake or outlet filter; or

replace fan as needed.

Correct any leaks.

Replace exhaust filter.

Replace compressor.

Replace absorbent beds.

Replace manifold or solenoid

valve.

Unit does not deliver

oxygen.

Cannula kinked or not properly

connected.

FreeStyle 5 is not turned on.

Battery is dead.

Defective AC or DC
power supply.

OCD sensor defective.

OCD solenoid valve defective.

Check cannula and the

cannula connection.

Turn FreeStyle 5 on to the

prescribed setting.

Replace battery.

Replace AC or DC power
supply.

Replace FreeStyle 5
main circuit board.

Replace product manifold or

OCD defective valve.

Unit continuously flows
oxygen (instead of pulse
flow delivery).

Defective OCD valve and

volume tank.

.

Replace product manifold.