Problem, Possible cause, Solution – AirSep MN179-1 B User Manual
Page 56

MN179-1 _ 02/14 AirSep Corp. - FreeStyle 5 Service Manual
Rev B
Problem
Possible Cause
Solution
Oxygen concentration
is below specifications
in test mode.
Unit’s temperature is too
high. Blocked air intake or
outlet; or defective fan.
Leak
Restriction in exhaust
felt.
Compressor has reached its
expected life.
Contaminated absorbent beds.
Defective solenoid valve.
Clear any obstruction in
air intake or outlet filter; or
replace fan as needed.
Correct any leaks.
Replace exhaust filter.
Replace compressor.
Replace absorbent beds.
Replace manifold or solenoid
valve.
Unit does not deliver
oxygen.
Cannula kinked or not properly
connected.
FreeStyle 5 is not turned on.
Battery is dead.
Defective AC or DC
power supply.
OCD sensor defective.
OCD solenoid valve defective.
Check cannula and the
cannula connection.
Turn FreeStyle 5 on to the
prescribed setting.
Replace battery.
Replace AC or DC power
supply.
Replace FreeStyle 5
main circuit board.
Replace product manifold or
OCD defective valve.
Unit continuously flows
oxygen (instead of pulse
flow delivery).
Defective OCD valve and
volume tank.
.
Replace product manifold.