Rev B
Figure 5.25
4. Loosen and remove the center bolt from the PSA assembly using a 5/32”
Allen wrench.
5. Holding sieve beds in place, gently pull upward on the product manifold to remove.
Keep tubing assembly in place.
6. Align new product manifold with the outlet of the sieve beds, and gently push
manifold down into place.
7. Reconnect tubing to new product manifold.
8. Place center bolt through product manifold and tighten to 15 in-lbs.
See also other documents in the category AirSep Health and hygiene:
- MN170-1 D (40 pages)
- MN172-1 (502 pages)
- MN189-1 (12 pages)
- MN123-1 F (45 pages)
- MN131-1 B (556 pages)
- MN166-1 B (40 pages)
- MN167-1 (520 pages)
- MN168-1 (3 pages)
- MN169-1 (40 pages)
- MN092-1 D (13 pages)
- MN104-1 D (40 pages)
- MN112-1 (208 pages)
- MN119-1 D (4 pages)
- MN120-1 (44 pages)
- MN078-1 F (42 pages)
- MN084-1 D (137 pages)
- MN139-1 (144 pages)
- MN164-1 (30 pages)
- MN152-1 (18 pages)
- MN136-1 H (27 pages)
- MN137-1 D (190 pages)
- MN156-1 B (18 pages)
- MN157-1 B (332 pages)
- MN165-1 G (28 pages)
- MN175-1 (2 pages)
- 15062781 (16 pages)
- 4111-7 (95 pages)
- 20626704 (24 pages)
- 20626704-C1 (154 pages)
- ML-CONC0056 (2 pages)
- MN028-1 (1 page)
- MN171-1 (23 pages)
- MN083-1 (1 page)
- MN126-1 C (69 pages)
- MN110-1 (30 pages)
- MN105-1 C (52 pages)
- MN118-1 B (40 pages)
- MN134-1 B (40 pages)
- MN151-1 (1 page)
- MN188-1 (1 page)
- MN188-2 (2 pages)
- MN155-1 (16 pages)
- MN138-1 H (46 pages)
- MN178-1 (1 page)