Appendix b: static calibration – Sterlco SF1000 Gravimetric Additive Feeder User Manual

Page 33

background image

33

Appendix B: Static calibration


1. General

- Static calibration is composed of two sub-procedures: ZERO calibration and GAIN

calibration.

- Special equipment required: A 1000-2500 gr. standard weight.

- Performing Static Calibration requires opening a password

2. Opening Password


a) While in Main Screen press the

key

"Maintenance” screen is displayed

b) Press , "Open/Close Password" screen

is displayed

c) Select (by

keys) "Open Password" and

Press

– "Enter Password" screen is

displayed

d) Enter Password "11" (by

keys) and

press

"Password Opened" screen is displayed

e) If the wrong password is entered' "Wrong pass"

screen will appear. Press to return to

"Open/Close Password" screen and try again

f) Once "Password Opened" screen appears, press twice to return to the

Maintenance screen (the password remains open!)

As long as the Password is open

– Static Calibration is possible