beautypg.com

Selecting the pulse output frequency mode, Operating procedures – Dwyer Series PUB User Manual

Page 31

background image

3: Operating Procedures

27

Note: This does not apply if the

PULSE OUTPUT menu is accessed by pressing the Pulse key when operating in

the

FLOW READING mode.

Calibrating the pulse output signal range (volumetric mode)

Note: The pulse

Output must be Off in order to change the Volume per pulse setting.

4.

Select a Pulse width (in ms) to suit the particular application – e.g. electro-mechanical counter. Refer to the
manufacturer’s data sheet for the minimum pulse width.

5.

Select

Exit and press the ENTER key to return to the FLOW READING screen.

Selecting the pulse output frequency mode

1.

With the instrument operating in the

FLOW

READING mode, press the Pulse function key to
access the

PULSE OUTPUT screen.

2.

To change the

Flow units shown in this menu

you must return to the

FLOW READING screen

and select the required units using keys 7,8 and
9.

PULSE OUTPUT

DD-MM-YY HH:MM:SS

Pulse output is ON

Qxx.xx%

Flow units : gal

Output : On

Vol per pulse : 10.00

Pulse width (ms) : 10

Max Pulse Freq (Hz) : 10.00

Flow at Max Freq : 200.00

3.

Select

Vol per pulse and enter the required

value.
In the example shown, a 10 ms pulse is
produced for every 10 gallons of flow.

Calculated Pulse Value: 20.0

Exit

1.

With the instrument operating in the

FLOW

READING mode, press the Pulse function key to
access the

PULSE OUTPUT screen.

2.

To change the pulse output to ‘frequency’ mode,
select the

Output menu option then select

Frequency as required.

3.

A1 Pulse Frequency is ON message
appears in the second line of the display.

4.

Select

Exit and press the ENTER key to return

to the

FLOW READING screen.

PULSE OUTPUT

DD-MM-YY HH:MM:SS

A1 Pulse Frequency is ON

Qxx.xx%

Flow units : gal

Output : Frequency

Vol per pulse : 10.00

Pulse width (ms) : 10

Max Pulse Freq (Hz) : 10.00

Flow at Max Freq : 200.00

Calculated Pulse Value: 20.0

Exit