beautypg.com

20ma signal scaling, Operating procedures – Dwyer Series PUB User Manual

Page 29

background image

3: Operating Procedures

25

Signal calibration

4-20mA Signal scaling

Note: The 4 to 20 mA output can be set to represent a particular flow range. It is also possible to enter a negative
figure for the minimum output and this would enable a reverse flow to be monitored.

10. Select

Output mA for error and enter a value (default is 22 mA) that you want the 4 to 20 mA output to

produce in the event of an error (e.g. if the flow-rate is outside the set range).

11. Upon completion, press the

ENTER key to return to the FLOW READING screen.

1.

Select

Setup Instrument from the MAIN

MENU, to access the SETUP INSTRUMENT
screen.

2.

Select

Calibrate 4-20mA.

SETUP INSTRUMENT

DD-MM-YY HH:MM:SS

Qxx.xx%

Set Date & Time : dd-mm-yy hh:mm:ss

Calibrate 4-20mA

Pulse output

Backlight : Disabled

Factory settings

Change Language

Exit

3.

Connect a calibrated ammeter to the 4 to 20 mA
output and adjust the

UP/DOWN Scroll keys

(Coarse) and

LEFT/RIGHT Scroll keys 5 & 6

(fine) until the output is exactly 4.0 mA.
The DAC should indicate approximately 8000.

4.

Press the

ENTER key when done.

CALIBRATE 4mA

DD-MM-YY HH:MM:SS

Qxx.xx%

Adjust the output current to 4mA

Use

UV

to set, 5/6 to trim

DAC Value: 8000

Press

when done

5.

With the meter still connected to the 4 to 20 mA
output adjust the Scroll keys to obtain an output
of exactly 20 mA.
The DAC should indicate approximately 40000.

6.

Press the

ENTER key when done.

CALIBRATE 20mA

DD-MM-YY HH:MM:SS

Qxx.xx%

Adjust the output current to 20mA

Use

UV

to set, 5/6 to trim

DAC Value: 40000

Press

when done

7.

With the instrument operating in the

FLOW

READING mode, press the 4-20mA function key.
This will access the

4-20mA OUTPUT screen.

8.

Select

Flow at max. output and enter a

value of the flow rate that you want to associate
with a 20 mA output.

9.

Select

Flow at min. output and enter a

value of the flow rate that you want to associate
with a 4 mA output. This could be ‘0’.

4-20 mA OUTPUT

DD-MM-YY HH:MM:SS

4-20 mA O/P is ON

Qxx.xx%

mA Output Reading : 0.00

Output Range : 4-20

Units : gal/min

Flow at max. output : 0.00

Flow at min. output : 0.00

Output mA for error : 22.00

Exit