beautypg.com

Attaching and connecting the transducers, Taking a flow reading, Flow monitoring – Dwyer Series PUB User Manual

Page 21: Operating procedures

background image

3: Operating Procedures

17

Note: If you press the ENTER key before the transducers are fitted and connected to the instrument, the
instrument will detect a low signal strength and may indicate an ERROR condition.

Attaching and connecting the transducers

14. Fit the designated sensors to the pipe using the appropriate transducer holders as described in Paragraph

2.2. Take great care to set the separation distance as accurately as possible.

15. Connect the red and blue coaxial cables between the sensors and the instrument, ensuring that the red

connector on the instrument is connected to the ‘upstream’ sensor.

Taking a flow reading

Flow monitoring

The

FLOW READING screen is the one most used during normal monitoring operation. It shows the instantaneous

fluid flow together with totalized values (when enabled). In this mode you can select the flow rate measurement
units by pressing keys 7 (liters), 8 (Gallons, Barrels) or 9 (m³), or change the display to show velocity by pressing
key 4.

11. The

SENSOR SEPARATION screen now displays

a summary of the entered parameters and
informs you of the type of sensor to be used, the
mode of operation, and the distance to set up
between the sensors.
In this example, it recommends type

A-ST (A

standard) sensors operating in the ‘

Reflex’

mode spaced

1.7 inches apart.

SENSOR SEPARATION

DD-MM-YY HH:MM:SS

Site : MySite

Pipe : 4.00 Inches

Wall : 0.20

Sensors : A-ST Reflex

Temperature : 5.00°C 41.00°F

Set sensor separation to 1.7 Inches
Press

to continue,

UV

to select sens.

12. Take a note of the displayed parameters, then

press the

ENTER key.

13. The

ATTACH SENSORS screen displays, giving

instructions to attach the sensors.

ATTACH SENSORS

DD-MM-YY HH:MM:SS

Attach Sensor Set

Red Connector UpStream

Press

key to go to Flow Reading

16. Once the transducers have been fitted and

connected, press the

ENTER key from the

ATTACH SENSORS screen.

17. This will take you to the

FLOW READING screen

via a signal-checking screen (shown here).

Please wait..

Checking signals

****************************************

* *

****************************************

18. Check that the indicated signal strength on the

left of the screen is at least 2 bars (ideally 3 or 4).
If less than 2 bars are shown it indicates there
could be a problem with the transducer spacing,
alignment or connections; or it could be due to
an application problem.

FLOW READING

DD-MM-YY HH:MM:SS

Qxx.xx%

Signal

gal/min

+Total: 0.00 gal

–Total: 0.00 gal

0.000