Operating procedures – Dwyer UFB User Manual
Page 37

3: Operating Procedures
33
3.
Select
Alarm 1 Mode and press ENTER to
select the
ALARM1 MODE menu shown below:
4.
Scroll down to
Frequency and press ENTER.
5.
This returns to the
ALARM SETTINGS menu
which will indicate
Frequency on the Alarm 1
Mode field as shown below.
6.
Select
Exit to return to the FLOW READING
OPTIONS menu.
7.
Scroll down to select
Max Pulse Freq.
(Hz) and press ENTER.
8.
Enter the required maximum pulse frequency,
e.g.
10Hz or 200Hz and press ENTER.
9.
Select
Flow at Max Frequency and press
ENTER.
10. Enter the maximum flow rate (in liters per
second) corresponding to 10Hz (or 200Hz) and
press
ENTER.
11. The
Calculated Pulse Value indicates
the relationship between the frequency and the
flowrate. This is a non-editable value.
12. Select
Exit and press ENTER to return to the
FLOW READING screen.
The message ‘
Frequency Pulse is ON’
should now be displayed on the status line of
the display (line 2).
Note: ALARM 2 can be used instead of ALARM 1. The procedure is identical except that
Alarm 2 Mode is
selected and the frequency pulse output is connected to ALARM2+ and ALARM2-.
ALARM SETTINGS
* *
Alarm1 Mode Off
Alarm1 Level :
Alarm2 Mode Off
Alarm2 Level :
Exit
ALARM1 MODE
* *
Off
Low
High
Volume
On Flow Error
Alarm Test
Frequency
ALARM SETTINGS
* *
Alarm1 Mode Frequency
Alarm1 Level :
Alarm2 Mode Off
Alarm2 Level :
Exit
FLOW READING OPTION
* *
Data review
Zero Cutoff (m/s) : 0.010
Set zero flow (m/s) : 0.000
Damping (secs) : 10
Totalizer : Run
Reset +Total
Reset –Total
Calibration factor : 1.000
Roughness factor : 0.010
Alarm Settings :
Max Pulse Freq (Hz) : 10.00
Flow at Max Frequency : 200.00
Calculated Pulse Value: 20.0
Diagnostics
Exit