Page 21: Instrument setup, Paragraph 3.2, Operating procedures
3: Operating Procedures
17
3.2
Using the Quick Start Menu
The Quick Start menu gathers various data for the site to be monitored and returns details of the transducer
configuration that must be applied when mounting the transducers on the pipe.
Before you can use the UFB system you need to obtain the following details (this information is required when
setting up the
Quick Start menu):
• The pipe outside diameter.
• The pipe wall thickness and material.
• The pipe lining thickness and material (if any).
• The type of fluid contained in the pipe being monitored.
• The fluid temperature.
Entering the site data
1.
Select
Quick Start from the MAIN MENU and press ENTER. You will then be presented with a series of
screens in which to enter the data mentioned above.
2.
Select the dimension units (millimeters or
inches) used to measure the pipe, then press
ENTER.
3.
Enter the pipe outside diameter dimension,
then press
ENTER.
4.
Enter the pipe wall thickness dimension, then
press
ENTER.
5.
If the pipe has a lining, enter the lining
thickness.
If nothing is entered the instrument
automatically assumes there is no lining.
6.
Press
ENTER to continue.
DIMENSION UNIT
* *
Select the dimension units:
mm
➥
Inches
OUTSIDE DIAMETER
* *
Dimensions: Inches
Pipe outside diameter: 6.50
PIPE WALL THICKNESS
* *
Dimensions: Inches
Pipe outside diameter: 6.50
Pipe wall thickness: 0.50
PIPE LINING THICKNESS
* *
Dimensions: Inches
Pipe outside diameter: 6.50
Pipe wall thickness: 0.50
Pipe lining thickness: 0.12