3: Operating Procedures

22

3.3.3 Adjusting the calibration factor

With the system running:

1.

Stop (

Stall) the totalizer facility and zero it (Paragraph 3.5).

2.

Run the totalizer to measure the total flow over a 30-60 minute period, and note the total flow indicated by the
reference flow meter over the same period.

3.

Calculate the % error between the UFB instrument and reference meters. If the error is greater than ±1%
calibrate the UFB as detailed below.

4.

Press the

Options key to access the FLOW

READING OPTION screen shown.

5.

Scroll down and select

Calibration factor

then press

ENTER.

6.

Change the calibration factor according to the
error calculated in step 3. For example, if the
instrument was reading 1% high then increase
the

Calibration factor value by 0.010.

Conversely, if the reading is 1% low then
decrease the calibration factor to 0.990.

7.

Press

ENTER to apply the change.

8.

Select

Roughness factor or Exit as

required and press

ENTER.

Key Point: In order to cancel any applied offset, you must read the flow via Quick
Start. Any value that you trim-out using the offset adjustment will be added/
subtracted from the flow reading across the whole range.

Key Point: USE THIS FACILITY WITH CARE & ONLY WHERE NECESSARY
The instrument is fully calibrated before leaving the factory and under normal cir-
cumstances does not require further calibration when used on site.
This facility can be used to correct the flow indication where unavoidable errors occur
due to the lack of a straight pipe or where the sensors are forced to be fitted close to
the pipe-end, valve, junction etc.
Any adjustment must be made using a reference flowmeter fitted in the system.

FLOW READING OPTION

* *

Data review

Zero Cutoff (m/s) : 0.010

Set zero flow (m/s) : 0.000

Damping (secs) : 10

Totalizer : Run

Reset +Total

Reset –Total

Calibration factor : 1.000

Roughness factor : 0.010

Alarm Settings :

Max Pulse Freq (Hz) : 10.00

Flow at Max Frequency : 200.00

Calculated Pulse Value: 2.00

Diagnostics

Select Totals : +Total

Exit