beautypg.com

4 calibration – Thermo Fisher Scientific CyberScan DO 300 User Manual

Page 9

background image

Instruction Manual

DO 300

6

4 CALIBRATION

4.1

Preparing the meter for calibration

The amount of oxygen dissolved in water will depend on its temperature, atmospheric

pressure and its salinity. While the pressure and salinity values are manually entered into the

instrument, the temperature is being measured by the probe. It is therefore very important that

the temperature is calibrated if necessary prior to the DO calibration.

The measurements of % Saturation of DO will linearly affect the measurement for DO in mg/L.

Hence calibration in % Saturation of DO should be carried out first.

Before calibrating your meter, make sure to rinse the probe well with DI (de-ionised) water or

rinse solution.

NOTE: Do not let membrane surface of the probe

touch any other surface. The probe guard (the piece

with holes fitted over the end of the probe) protects the

membrane; make sure this is always attached to the

probe while it is in use.

You can calibrate for dissolved oxygen in either %

saturation or mg/L (ppm). When you recalibrate your

meter in % saturation mode, old calibrations in %

saturation will be replaced while the % saturation offset

adjustment will be erased. However, recalibration in the

mg/L (ppm) concentration mode will only replace the old

calibration in the mg/L (ppm) mode and do not affect the calibration in the % saturation mode.

Before Calibration

Before starting calibration, make sure you are in the correct measurement mode. When you

switch on the meter, the meter starts up in the mode last used (either in mg/L (ppm) or %

Saturation measurement mode or clock). For example, if you switch the meter off while it is in

the “mg/L” measurement mode, it will also be in the “mg/L” measurement mode when you

switch the meter on.

Figure 5: DO Probe guard